ClearanceBrief
Recall Z-1338-2020 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Luminos Agile Max, model no. 10762471 - Product Usage: Luminos Agile Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. The system has medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examinations of the urogenital tract.

DateJan 8, 2020
StatusTerminated
ReasonA software bug may lead to one image to be assigned to two different patients, which could potentially affect medical diagnosis.
Root causeSoftware design
ActionOn January 8, 2020, the firm notified customers of the product issue via an Urgent Medical Device Correction letter. Customers were instructed to ensure that the image displayed after acquisition is a newly acquired image and belongs to the current patient. If the error occurs and a previously acquired image is duplicated, the user should delete the image and contact Siemens Customer Service at 1-800-888-7436 to report the incident. Siemens Healthineers will correct the error with a field software update which will be available beginning January 2020. Your local Siemens Healthineers Service
Quantity475
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173639Luminos dRF Max, Luminos Agile Max, Uroskop Omnia Max, Multitom Rax
Siemens Medi Cal Solutions, Inc., Apr 2, 2018