Recall Z-1332-2020 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Luminos Agile Max, Model No. 10762472 - Product Usage: Luminos Agile Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image. The system has medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examinations of the urogenital tract.

DateJan 8, 2020
StatusTerminated
ReasonSiemens became aware of an issue of incorrect use of the bypass key overriding the built-in collision control creating risk of collision with obstacles or persons when the operator uses the override function.
Root causeUse error
ActionOn January 8, 2020, the firm initiated the recall, notifying customers via Urgent Medical Device Correction letters. Customers were advised of the potential for a system collision due to incorrect use of the bypass key overriding the built-in collision control. The bypass key is intended to enable movement of an otherwise blocked system to avert immediate hazard, e.g. to recover a patient following device error or in case of emergency. It is not intended for permanent use. Customers were instructed to use the bypass key exclusively to avert an immediate hazard and immediately contact Siemens
Quantity475
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173639Luminos dRF Max, Luminos Agile Max, Uroskop Omnia Max, Multitom Rax
Siemens Medi Cal Solutions, Inc., Apr 2, 2018