ClearanceBrief
Recall Z-1264-2020 ยท GDW Staple, Implantable

COVIDIEN LLC

Covidien Endo GIA Ultra Universal Stapler 12 mm-intended for resection, transection in general surgery Item Code: EGIAUSTND

DateJan 17, 2020
StatusTerminated
ReasonPotential for the sterile barrier to be compromised. Products with a compromised sterile barrier may result in a potentially increased risk for infection
Root causePackage design/selection
ActionMedtronic notified consignees by letter via Federal Express and/or certified mail beginning January 17, 2020. The letter informs of the reason for recall, health risk and action to take: Immediately quarantine and discontinue use of the affected itemcode and lot; " Return affected product; Forward the information to anyone who they have distributed. Complete the Recalled Product Return Form even if you do not have inventory. Questions: Email Medtronic Post Market Vigilance at: rs.assurancequality@medtronic.com.
Quantity2,180 units
DistributionUS Nationwide, Israel, and Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111825ENDO GIA(TM) STAPLERS, DST SERIES(TM) GIA(TM) STAPLERS, DST SERIES(TM) TA(TM) STAPLERS
Covidien, Formerly US Surgical A Divison of Tyco H, Oct 25, 2011