Recall Z-1375-2020 ยท NHA Abutment, Implant, Dental, Endosseous

Biomet 3i, LLC

CERTAIN BELLATEK ENCODE HEALING ABUTMENT 4.1MM(D) X 5MM(P) X 3MM(H) Reference Number: IEHA453

DateJan 10, 2020
StatusTerminated
ReasonResulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin contour misalignment
Root causeProcess control
ActionZimmer Biomet issued letter to Distributors, Clinicians via FedX on 1/10/20 stating reason for recall, health risk and action take: Step 1: Quarantine any unused affected products in your inventory for return to Zimmer Biomet Dental. Step 2: Identify any patients currently in your care who have received an affected Certain Encode Healing Abutment with the lot numbers in Table 1. a)If the lot number of the Certain BellaTek Encode Healing Abutment is known: a.If the patient has not yet undergone scanning/impressioning, the affected Certain BellaTek Encode Healing Abutment should be replaced wi
DistributionNationwide Foreign: Australia Austria Belgium China France Germany Italy Japan Netherlands Poland Spain UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072642BIOMET 3I DENTAL ABUTMENTS AND RESTORATIVE COMPONENTS
Biomet 3i, Inc., Dec 20, 2007