ClearanceBrief
Recall Z-1277-2020 ยท GAM Suture, Absorbable, Synthetic, Polyglycolic Acid

Covidien LLC

Polysorb Braided Absorbable Suture 3/0 30 VIOLET GS-11, Item Code CL815 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in cardiovascular or neural tissue.

DateFeb 19, 2020 (initiated Dec 18, 2019)
StatusTerminated
ReasonThere is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
Root causeEquipment maintenance
ActionOn December 18, 2109, Medtronic issued Urgent Medical Device Recall and Return Form notices to customers via courier service. Customers were advised to take the following required actions: 1. Please immediately quarantine and discontinue use of the affected item code and lot listed within the notice. 2. Please return affected product as indicated within the notice. All unused products from the affected item codes and lots must be returned. 3. If you have distributed any of the Polysorb braided or Biosyn monofilament absorbable sutures listed on the recall notice, please promptly forward
Quantity139,938 units total
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K000037MONOFILAMENT BIOSYN SUTURE
United States Surgical, A Division of Tyco Healthc, Mar 30, 2000