Recall Z-1717-2020 ยท KQO Automated Urinalysis System

Sysmex America, Inc.

Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.

DateJan 21, 2020
StatusTerminated
ReasonWhen excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.
Root causeProcess control
ActionThe firm initiated the recall by letter on January 21, 2020. The "Product Notification", identified the Field Correction and instructed the customer to use the kit to direct fluid safety away from the conveyor electrical parts to address this potential hazard. During the next service visit, this kit will be installed by your Sysmex Field Service Engineer. The Notification also instructs the customer to: 1. Ensure that the surface of the sample, in the two, is more than away from the opening of the two. 2. Do not store samples on or above the analyzer. 3. if the sample transport convey
Quantity40 units
DistributionUSA: FL, VT, IL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.