Recall Z-1233-2020 ยท JFY Enzymatic Method, Creatinine

Siemens Healthcare Diagnostics, Inc.

Dimension Vista Enzymatic Creatinine (ECREA) reagent, SMN 10700444

DateJan 24, 2020 (initiated Dec 18, 2019)
StatusTerminated
ReasonThere is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.
Root causeUnder Investigation by firm
ActionSiemens Healthcare Diagnostics initiated a customer communication for this issue via letter beginning December 18, 2019.
Quantity11,725 units
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090330DIMENSION VISTA ENZYMATIC CREATININE FLEX REAGENT CARTRIDGE, AND ENZYMATIC CREATININE CALIBRATOR, MODELS K1270, KC270
Siemens Healthcare Diagnostics, Inc., May 22, 2009