ClearanceBrief
Recall Z-1018-2020 ยท FYA Gown, Surgical

Cardinal Health

AAMI 3 NONREIN SURG GOWN L BNS, Item Code 9515NB

DateJan 24, 2020 (initiated Jan 11, 2020)
StatusTerminated
ReasonThe gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
Root causeVendor change control
ActionA letter dated 1/11/20 informed customers that: Cardinal Health has become aware of a quality issue related to the manufacture of certain AAMI Level 3 surgical gowns. We have initiated an investigation, placed a hold on product inventory, and are working with the U.S. Food and Drug Administration (FDA) to address this issue. This inventory hold affects certain lots of bulk non-sterile and single-sterile Cardinal Health" surgical gowns and Presource procedure packs that include these surgical gowns. Your local sales representative will provide specific details. Cardinal Health intends to init
Quantity7,659,634 total units
DistributionWorldwide distribution US Nationwide including Puerto Rico and Guam, United Arab Emirates, Australia, Canada, Chile, Colombia, Costa Rica, Dominican Republic, Israel, Mexico, Oman, Panama, Philippines, and Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170762Cardinal Health Non-Reinforced Surgical Gown
Cardinal Health200, LLC, Jun 9, 2017