Recall Z-1563-2020 ยท LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy

LivaNova USA Inc

VNS Therapy Programmer, Model 3000 used with Models 103, 104, 105, 106, or 8103 VNS Therapy generators.

DateFeb 4, 2020
StatusOpen, Classified
ReasonFalse positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3) In-session re-interrogation performed. Users instructed to lower output current and widen pulse width. Only system diagnostic testing evaluates output current. Users may conclude device malfunction, could lead to medical/surgical intervention.
Root causeSoftware Design Change
ActionOn 02/04/20, Urgent Medical Device Correction notices were mailed to Doctors, who were informed the following: Which Patients are Potentially Impacted? Because the false positive warning message only appears if the generator is initially interrogated at 0mA, there are only certain scenarios where the issue may appear: - After initial implant, when the device is programmed on for the first time; - Following an MRI, after the device was programmed to 0mA per labeling instructions; or - Resuming therapy following previous device disablement due to an adverse event or device issue. Actions to be
Quantity2683
DistributionU.S.: LA, ID, AZ, UT, KY, VA, TN, NY, SC, IL, IN, NC, WA, CA, OH, FL, PA, NJ, CO, TX, RI, MD, HI, OR, MA, NE, MO, OK, MS, MN, SD, VT, AR, WV, WI, MI, AL, KS, DE, NM, GA, ND, AK, MT, ME, NV, DC, MO, WY, IA, CT. Canada, and Israel.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.