ClearanceBrief
Recall Z-1262-2020 ยท JGS Electrode, Ion Specific, Sodium

Abbott Point Of Care Inc.

i-STAT CHEM8+ cartridges (blue), List No. 09P31-25. UDI (01) 00054749001910 - Product Usage: The i-STAT CHEM8+ cartridge may be used for the quantitative measurement of sodium, potassium, chloride, blood urea nitrogen (BUN), ionized calcium, TCO2, glucose, creatinine, and hematocrit in venous, arterial or capillary whole blood.

DateFeb 18, 2020 (initiated Jan 15, 2020)
StatusTerminated
ReasonThe I-STAT BLUE CHEM8+ and CG4+ cartridges are not FDA cleared and do not have CLIA waived status. CLIA recognizes two types of laboratory tests: waived and non-waived. Different CLIA standards apply to waived and non-waived tests as well as the facilities that perform these types of tests. The i-STAT CHEM8+ cartridge contains nine measured assays (sodium, potassium, chloride, blood urea nitrogen (BUN), ionized calcium, TCO2, glucose, creatinine and hematocrit) and is categorized as a CLIA waived test for venous whole blood samples, meaning this cartridge could be used in a facility with a CL
Root causeNo Marketing Application
ActionOn 1/15/2020, the firm notified customers of the product recall via letters, customer emails, follow-up phone calls/visits by sales representatives, and a website posting at www.pointofcare.abbott. Customers were instructed to do the following: Facilities using the i-STAT BLUE CHEM8+ cartridges in CLIA waived settings should transition to alternate CLIA waived testing methods or engage a reference laboratory for the tests your facility requires. Facilities should discontinue use of the i-STAT BLUE CHEM8+ and CG4+ cartridges with capillary samples. Facilities should use an alternate method,
Quantity9,049,725
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.