Recall Z-1118-2020 ยท LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy
LivaNova USA Inc
VNS Therapy SENTIVA Generator Model # 1000 Labeling 26-0009-6800 Model 1000 Generator Sample Labels 26-0009-9600 Patient s Manual for Depression 26-0009-9500 Patient s Manual for Epilepsy 76-0000-5000 VNS Therapy System Physician s Manual (refer to Attachment 1)
| Date | Aug 13, 2019 |
| Status | Terminated |
| Reason | Firm identified a subset of its generators that were sterilized one additional sterilization cycle, which does not meet the firm's quality specifications. |
| Root cause | Process control |
| Action | On 08/13/2019, the firm contacted via telephone hospitals/facilities of the immediate need to remove and replace the affected devices. Reference Firm email communication dated 1/27/2020. On 08/30/2019, prescribing physicians were telephone call with a follow-up email recommending replacement of the generators during an already planned or unrelated lead replacement surgery. |
| Quantity | 6 generators |
| Distribution | US: FL, GA, KS, MO, and UT. No OUS distribution |
| FDA record | Open on fda.gov |