Recall Z-1145-2020 ยท OWB Interventional Fluoroscopic X-Ray System

Philips North America, LLC

VesselNavigator application used with Philips Azurion (Azurion 7 Series) Software R1.2 and R2.0, and Philips Allura Xper R9, upgraded to Azurion 1.2 M20. System codes: Azurion 7 B12 (722067), Azurion 3 B20 (722068), Azurion 7 M12 (722078), Azurion 7 M20 (722079), Field Extensions Azurion 7 series (722136), Allura Xper R9 (722303). The product is software, used during an endovascular interventio

DateFeb 10, 2020 (initiated Dec 23, 2019)
StatusTerminated
ReasonDue to a software defect, when a digital subtraction angiography (DSA) is exported to the VesselNavigator application, the DSA is displayed without subtraction.
Root causeSoftware design
ActionOn 12/23/2019, the firm sent medical device correction letters to affected customers via certified mail. The firm will be contacting customers with affected systems. The firm will be implementing software updates in all affected systems to correct the issue which was made available by the end of December 2019. The user was asked to do the following: -Please suspend the use of Digital Subtraction Angiography images for the verification of the registration during the Live Guidance step. -Please ensure that all staff working with the VesselNavigator application are informed of the recall. If y
Quantity443
DistributionWorldwide Distribution. US nationwide, Canada, Argentina, Australia, Austria, Belgium, Brazil, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, France, Germany, Greece, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Latvia, Libya, Mexico, Netherland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162148Allura Xper R9
Philips Medical Systems Nethrlands BV, Nov 23, 2016
K172822Azurion series R1.2
Philips Medical Systems Nederland B.V., Nov 22, 2017
K181830Azurion R2.0
Philips Medical Systems Nederland B.V., Aug 2, 2018