Recall Z-0860-2020 ยท GAA Needle, Aspiration And Injection, Disposable

Vitreq Bv

VitreQ 25G Blunt Needles with luer-lock connector-REF : CN25.D05 GTIN: 8719214221645

DateDec 17, 2019
StatusTerminated
ReasonMicroscopic tears of the sterile pouch may compromise sterility
Root causePackage design/selection
ActionVitreQ issued Urgent Medical Device recall dated 12/16/19 letter states reason for recall, health risk and action to take: 1.Immediately examine your inventory and quarantine product from all lots subject to recall. If you have further distributed this product, please identify your customers and notify them at once of this product recall. Consider all potential users of this product in your user supply chain. You may provide a copy of this recall notification letter; 2.Complete the Attachment 1: Response Form enclosed IMMEDIATELY, as evidence of the product being returned, and we will credit y
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.