ClearanceBrief
Recall Z-0807-2020 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Siemens ARTIS pheno, Model # 10849000. Interventional Fluoroscopic X-Ray System - Product Usage: Angiography system for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

DateDec 20, 2019
StatusTerminated
ReasonThe potential exists for system movement to be permanently blocked by an activation of the collision sensor integrated in the cover of the collimator even when no collision has occurred.
Root causeDevice Design
ActionSiemens Medical Solutions USA, Inc. sent a Customer Safety Advisory Notice to all affected customers via AX070/19/S on 12/20/2019.
Quantity6
DistributionUS Nationwide distribution in the states of FL, IL, IN, MA, PA, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163286ARTIS pheno
Siemens Medical Solution USA, Inc., Mar 9, 2017