Recall Z-2105-2020 ยท LZH Pump, Infusion, Enteral

Cardinal Health Inc.

Kangaroo 924 Safety Screw Spike Set, Product Code 775759

DateDec 26, 2019
StatusOpen, Classified
ReasonThe feeding spike sets may leak at the interface of the tube and spike connector.
Root causeProcess control
ActionA Medical Device Product Advisory letter dated 12/23/19 was sent to customers. 1) INSPECT your inventory for affected product. 2) COMMUNICATE with all personnel that utilize the equipment the risk of leaks at the interface of the tube and spike connector. 3) INSPECT the feeding set for leaks after the pump has been primed and the infusion has started. " If a leak is detected the feeding set must be discarded and a new feeding set must be obtained. Inspect the new feeding set for leaks in the same manner. 4) NOTIFY any customers to whom you may have distributed or forwarded affected product ab
Quantity114,731, 290 total units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141479KANGAROO ENTERAL FEEDING SET WITH ENFIT CONNECTORS
Covidien, Dec 23, 2014