ClearanceBrief
Recall Z-1124-2020 ยท LKK Pump, Infusion, Implanted, Programmable

Flowonix Medical Inc

Prometra Programmable 20mL Pump. Catalog No. 91827 The device is an implantable infusion pump. GTIN-14 00810335020XXX (XXX = in-country language specific)

DateFeb 10, 2020 (initiated Dec 24, 2019)
StatusOpen, Classified
ReasonA pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
Root causeSoftware change control
Action***Updated 5/12/2023*** On May 11, 2023, the firm sent an URGENT MEDICAL DEVICE CORRECTION UPDATE letter to all affected customers, which informed them of an update to Flowonix's plan to correct the firmware issue. Flowonix will NOT be conducting an office-based pump firmware update for affected pumps implanted prior to January 2020. Healthcare providers were provided ongoing recommendations for Prometra and Prometra II pumps implanted prior to January 2020, which are impacted by this issue. Healthcare providers should contact their Flowonix Representative or our Technical Solutions Departm
Quantity1083 (OUS)
DistributionNationwide domestic distribution. Foreign distribution to United Kingdom, Republic of Ireland, Poland, Greece, Italy, Germany, Netherlands, and Belgium.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.