ClearanceBrief
Recall Z-1151-2020 ยท HAW Neurological Stereotaxic Instrument

MEDTECH SAS

ROSA One 3.1 Brain Application

DateJan 6, 2020
StatusTerminated
ReasonSome cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the surgery planning.
Root causeSoftware design
ActionOn January 6, 2020, Zimmer Biomet issued URGENT MEDICAL DEVICE CORRECTION notices to customers via FedEx. Actions Required By Users: Your device can continue to be used, as long as the recommendations below are followed: While the current Instructions For Use do not require the images to be reviewed with ROSA Brain software prior to surgery, they adequately describe the standard workflow and indicate that the image acquisition of the patients head (MRI / CT or 2D X-Ray images) is performed prior to surgery, and is then loaded into the device and used by the surgeon in the preoperative pha
Quantity39 units
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151359ROSA Brain
Medtech S.A, Dec 18, 2015
K172444ROSA BRAIN (v3.0.0.5)
Medtech S.A, Mar 28, 2018
K182417ROSA ONE Brain Application
Medtech S.A, Feb 7, 2019