Recall Z-1350-2020 ยท KWP Appliance, Fixation, Spinal Interlaminal

ulrich medical USA Inc

Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw

DateDec 31, 2019
StatusTerminated
ReasonDue to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.
Root causeProcess control
ActionUlrich notified initially notified customers, via phone call, on 12/20/2019. On about 12/31/2019, customers were also sent an official "URGENT SAFETY INFORMATION: MEDICAL DEVICE CORRECTIONS & REMOVALS" letter. Customers were informed that the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation. Instructions included cease use of the affected devices, return them to Ulrich, and complete and return the provided "Product Recall" form. Customers were also instructed to inform all users of the devices within each organization
Quantity179 units
DistributionAZ, TX, GA, FL, AR, CA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150650neon3
Ulrich GmbH & Co. KG, Sep 24, 2015