ClearanceBrief
Recall Z-0844-2020 ยท DQX Wire, Guide, Catheter

Arrow International Inc

Arrow Arterial Catherization Kit-20 Ga. x 4.45 cm Radiopaque Polyurethane over 22 Ga. TW Introducer Needle with Integral .018" (0.46 mm) dia. Spring-Wire Guide Product Code: ASK-04020-PMC The Arrow Arterial Catheterization device permits access to the peripheral arterial circulation or to other small vessels.

DateDec 12, 2019
StatusOpen, Classified
ReasonThe lidstock label incorrectly displays a MR (Magnetic Resonance ) Conditional symbol (instead of a MR Unsafe symbol) in reference to a component of the kit
Root causeLabeling Change Control
ActionArrow International issued an " Urgent Medical Device Recall Notice" to its US consignee on 12th Dec 2019 via FedEx 2-day mail. The letter states reason for recall, health risk and action to take: If you have affected stock in inventory, immediately discontinue use and quarantine the products. 2. Please complete the enclosed Acknowledgement Form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to recalls@teleflex.com. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instruction
Quantity80 eaches
DistributionUS distribution to SC
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K810675ARROW RADIAL ARTERY CATHETERIZATION SET
Arrow Intl., Inc., Mar 27, 1981