Recall Z-1001-2020 ยท GKZ Counter, Differential Cell

Clinical Diagnostic Solutions, Inc.

Medonic M-Series Hematology Analyzer M16S BD ABR US Product code: 1400075; OUS Product code 1400011

DateJan 2, 2020
StatusTerminated
ReasonA mix-up of autosampler tube positions with a possibility of misidentification of sample results and a risk of operator injury from an exposed aspiration needle.
Root causeSoftware design
ActionConsignees notified via email and postal delivery with recall letter dated 01/02/2020. The letter identified affected product, stated reason for recall, and provided specific actions to follow. A Recall Acknowledgement Form was requested to be returned by email, postal delivery, or fax. Clinical Diagnostic Solutions Inc. is handling US consignees only. Boule Medical AB is handling OUS consignees.
Quantity261
DistributionWorldwide distribution. US Nationwide, Venezuela, Africa, Italy, Brazil, and Ukraine. No governmental, Canadian, or Mexican consignees.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083420MEDONIC M SERIES
Boule Medical AB, Feb 3, 2009