Recall Z-0819-2020 ยท LLZ System, Image Processing, Radiological

GE Healthcare, LLC

Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from other healthcare information sources.

DateJan 17, 2020 (initiated Dec 16, 2019)
StatusTerminated
ReasonCentricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.
Root causeSoftware design
ActionThe firm initiated their correction on 12/16/2019 by letter. The letter explained the issue and directed the consignee to follow instructions in the letter until an update to the Centricity Universal Viewer can be made by the firm to correct the issue.
Quantity205
DistributionDistributed nationwide; Internationally distributed to Mexico and 24 additional countries.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150420Centricity Universal Viewer
GE Healthcare, May 26, 2015
K182419Centricity Universal Viewer
GE Healthcare, Dec 14, 2018