Recall Z-0815-2020 ยท GBO Catheter, Nephrostomy, General & Plastic Surgery

Cook Inc.

Dawson-Mueller Drainage Catheter, Multipurpose Drainage Catheter, RPN UL T6.3-35-15-P-5S-CLDM-HC, UL T6. 3-35-25-P-58-CLDM-HC, UL T7.0-35-25-P-5S-CLDM-HC

DateDec 27, 2019
StatusTerminated
ReasonSpecific lots of the Dawson-Mueller Drainage Catheter were not manufactured to specification, which could lead to leakage from the Mac-Loc hub assembly.
Root causeEmployee error
ActionOn December 27, 2019, the firm sent Urgent Medical Device Recall letters to affected customers, advising them of the recalled products. Customers were asked to take the following actions: 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of r
Quantity1604
DistributionNationwide domestic distribution. Foreign distribution to Australia, Belgium, Canada, China, Colombia, Denmark, France, Germany, Italy, Korea, Netherlands, New Zealand, Peru, Russia, Spain, Sweden, and United Kingdom.