Recall Z-0841-2020 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Palodex Group Oy

KaVo Scan eXam SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)

DateJan 7, 2020
StatusTerminated
ReasonSlow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.
Root causeComponent design/selection
ActionOn January 7, 2020 the firm distributed Urgent Medical Device Recall Communication letters by letter stating: Urgent Medical Device Correction Dear KaVo Kerr Customer, Palodex Group OY has identified an issue with Soredex DIGORA Optime, KaVo Scan eXam, Instrumentarium Express, and Gendex GXPS-500 Imaging Plate Systems (later called IPS). According to our records, we have delivered to you one or more devices belonging to the serial number range defined in the table below. The Power Supply Unit (PSU) delivered with the device may have a fault requiring correction. Product Soredex DIGORA Optime
Quantity4,563 Total Globally (all products)
DistributionUS: WI and PA OUS: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KOREA, REPUBLIC OF, KUWAIT, MALAYSIA, MALTA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041050DIGORA OPTIME (OR DIGORA ONTIME)
Soredex Palodex Group OY, Jun 15, 2004