Recall Z-1045-2020 ยท OFF Central Venous Catheter Tray
Centurion Medical Products Corporation
Centurion PICC : CVI2650 ANGIO TRAY CVI3610 CENTRAL LINE INSERTION BUNDLE CVI4025 PICC INSERTION BUNDLE (P/S) CVI4190 CENTRAL LINE INSERTION BUNDLE CVI4190 CENTRAL LINE INSERTION BUNDLE CVI4330 PICC INSERTION TRAY CVI4340 MIDLINE INSERTION TRAY CVI4485 MOUNT SINAI CVP INSERTION KIT CVI515 VENOUS ACCESS KIT CVI650 INSERTION TRAY
| Date | Dec 11, 2019 |
| Status | Terminated |
| Reason | Incomplete seals on the sterile package may compromise the sterility |
| Root cause | Packaging process control |
| Action | Centurion issued a notification emailed to Medline Industries, Centurion's parent company. Centurion issued mailing letters via First Class Mail beginning on 12/12/2019, letter states reason for recall, health risk and action to take: Please identify all inventory that you have within your possession and segregate it in a location where it cannot be accessed for use. 2) If you have transferred possession of this product to another individual, to another department or to another location in your organization, you are required by FDA regulations to notify them of this recall communication. Pleas |
| Quantity | 29020 |
| Distribution | Nationwide Foreign: GHANA |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.