Recall Z-1311-2020 ยท DZI Drill, Bone, Powered
Nobel Biocare Usa Llc
Drill Tapered WP 5x10 mm Single-Pat - Product Usage: Drills are intended to be used in the upper and lower jaw bone to prepare osteotomy prior to implant placement.
| Date | Feb 19, 2020 (initiated Mar 14, 2019) |
| Status | Terminated |
| Reason | Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life. |
| Root cause | Packaging process control |
| Action | On 3/27/2018 the firm issued a recall notice by email to affected consignees. The recall notification informed the following actions to be taken by the consignees: 1. Identify affected devices as detailed in the Customer Reply Form. 2. Do not use affected devices. 3. Send back the Customer Reply Form and the affected devices. The firm will replace affected devices free of charge. This letter needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. We kindly ask you to follow instru |
| Quantity | 264 units |
| Distribution | Worldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K990846 | AMORPHOUS DIAMOND COATED DRILL Nobel Biocare Uas, Inc., Jun 10, 1999 |