ClearanceBrief
Recall Z-1304-2020 ยท DZI Drill, Bone, Powered

Nobel Biocare Usa Llc

Dense Bone Drill 6.0 6x16 mm Single-Pat- Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants.

DateFeb 19, 2020 (initiated Mar 14, 2019)
StatusTerminated
ReasonDue to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
Root causePackaging process control
ActionOn 3/27/2018 the firm issued a recall notice by email to affected consignees. The recall notification informed the following actions to be taken by the consignees: 1. Identify affected devices as detailed in the Customer Reply Form. 2. Do not use affected devices. 3. Send back the Customer Reply Form and the affected devices. The firm will replace affected devices free of charge. This letter needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. We kindly ask you to follow instru
Quantity8 units
DistributionWorldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K990846AMORPHOUS DIAMOND COATED DRILL
Nobel Biocare Uas, Inc., Jun 10, 1999