ClearanceBrief
Recall Z-0837-2020 ยท FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days

Arrow International Inc

Arrow Endurance" Extended Dwell Peripheral Catheter System, REF: ASK-00820-FLH

DateDec 6, 2019
StatusTerminated
ReasonPotential sterility issue for specific lots of the Arrow Endurance" Extended Dwell Peripheral Catheter System, Product Code ASK-00820-FLH.
Root causeEmployee error
ActionOn December 6, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail stating: Urgent Medical Device Recall Notice Arrow Endurance" Extended Dwell Peripheral Catheter System EIF-000391 December 06, 2019 To: Customer of Arrow Products Arrow International is voluntarily recalling the following product code and lots: Product Code: ASK-00820-FLH Lot Numbers: 13F19E0247, 13F19F0006, 13F19G0433 Arrow International is recalling the above product code and lots due to a potential sterility issue. The product was processed through the incorrect sterilization cycle; thus,
Quantity575
DistributionUS: FL
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163513Arrow Endurance Extended Dwell Peripheral Catheter System
Arrow International, Inc. (Subsidiary of Teleflex, Inc.), Feb 24, 2017