Recall Z-0928-2020 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Biomerieux Inc

BIOMERIEUX VIDAS 3, Compact immunoanalyzer with full traceability and automation

DateOct 22, 2019
StatusTerminated
ReasonFollowing Customers complaints, investigations have been initiated on potentially false results obtained on VIDAS 3 with an expired calibration. Indeed, the calibrations of assays were valid in the calibration menu whereas in fact the calibrations were expired and no alarm displayed to warn the users about the expiration date of the calibrations. Software computes results of analysis for assays with expired calibrations. The anomaly is due to an incorrect update of the calibration status by the software.
Root causeSoftware design
ActionThe firm initiated their field action by an email to their subsidiaries on 10/22/2019. The subsidiaries then notified their customers by 2-day express via Fed Ex. In the US customer letter, customers are instructed to check their software version. If the VIDAS 3 software version is between 1.2.0 and 1.3.1, customers are instructed to check the archiving mode. If the archiving mode is Automatic When Results are Reported then customers are instructed how to configure the workaround, and to contact bioMerieux if the customer has the incorrect software configuration. bioMerieux then instructs
Quantity738 units
DistributionUS