ClearanceBrief
Recall Z-0813-2020 ยท BSZ Gas-Machine, Anesthesia

GE Healthcare, LLC

Carestation 620/650/650c A1 Product Usage: anesthesia systems are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (pediatric and adult.

DateJan 23, 2020 (initiated Nov 25, 2019)
StatusTerminated
ReasonGE Healthcare has become aware that there is a potential for a loose cable connection inside specific manufactured anesthesia devices. This would cause a loss of mechanical ventilation and the system will provide high priority audio and visual alarms. Loss of mechanical ventilation could lead to hypoxia if the clinician does not intervene.
Root causeProcess control
ActionThe firm initiated their field notification by letter on November 25, 2019. The letter states the reason for recall and provides the following safety instructions: You can continue to use the anesthesia system. - If you observe the message Ventilate Manually! , change from mechanical to manual ventilation. At any time, the clinician may use a self-inflating bag to ventilate the patient and/or switch to another anesthesia device. Contact your GE Healthcare representative for repair of the device. - Perform the planned maintenance (PM) every 12-months at a minimum per the User s Reference
Quantity3319 units
DistributionWorldwide Distribution: US (Nationwide) and countries of: Albania, Algeria, Argentina, Australia, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Croatia and Hrvatska, Czech Republic, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Guatem
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151570Carestation 620/650/650C
Datex-Ohmeda, Inc., Jan 21, 2016