Recall Z-0957-2020 ยท OZP Automated Insulin Dosing Device System, Single Hormonal Control

Medtronic Inc.

Medtronic MiniMed 620G Insulin Infusion Pump, Ref #s/Model(s) - MMT-1510, MMT-1710, and MMT-1750 *Not Distributed within the US

DateFeb 7, 2020 (initiated Nov 21, 2019)
StatusOpen, Classified
ReasonThere have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia.
Root causeDevice Design
ActionOn November 21, 2019, Medtronic initiated a recall of MiniMed 600 series insulin pumps with a clear retainer ring design. On October 5, 2021, the firm expanded the recall and issued an "Urgent Field Safety Notification" to all patients and health care providers. The updated notification informs consignees that Medtronic will proactively replace all MiniMed 600 series insulin pumps that have a clear retainer ring design with an equivalent pump that has the updated black retainer ring design. Pumps will be replaced at no charge over the coming months even if the retainer ring is not damaged and
DistributionUS:AK,AL,AR,AZ,BM,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY OUS: France, Germany, UK, Italy, Netherlands, Spain, Sweden, Israel, Norway, Austria, Belgium, Denmark, Czech Repub