Recall Z-0724-2020 ยท KNW Instrument, Biopsy

Bard Peripheral Vascular Inc

BARD MAX-CORE Disposable Core Biopsy Instrument - Product Usage: The core needle biopsy device is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone.

DateOct 25, 2019
StatusTerminated
ReasonDuring the firm's review of May 2019 complaint data, a significant increase in Disposable Core Biopsy Instrument complaints were observed for Failure to Fire, Firing Problem, and Failure to Obtain Samples.
Root causeEquipment maintenance
ActionOn 10/28/2019, the firm sent an "URGENT MEDICAL DEVICE RECALL NOTIFICATION" Letter to customers via FedEx informing them the firm has identified MAX-CORE Disposable Core Biopsy Instruments that may be at risk of having firing issues. An example of a firing issue is where the device does not completely fire, resulting in the inability to obtain tissue samples. The firm also informed their customers that using the affected product will lead to a varying degree of user dissatisfaction or may be associated with a prolonged procedure or minor tissue injury or minor bleeding. Most often, a coaxial
Quantity727,034 units
DistributionWorldwide distribution - US Nationwide distribution and countries of Australia, Canada, Europe, Middle East, Africa, Latin America, Asia Pacific.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133948BARD(R) MONOPTY(R) DISPOSABLE CORE BIOPSY INSTRUMENT, BARD(R) MAX-CORE(R) DISPOSABLE CORE INSTRUMENT
Bard Peripheral Vascular, Inc., Feb 21, 2014