Recall Z-0916-2020 ยท CCW Laryngoscope, Rigid

King Systems Corp. dba Ambu, Inc.

King Vision Video Adapter Size 1/2. Laryngoscope used to examine and visualize a patients upper airway and aid in the placement of a tracheal tube. Model Number: KVLVA12

DateFeb 4, 2020 (initiated Nov 5, 2019)
StatusTerminated
ReasonExhibiting a reversed image. Although the image may appear normal, the users actions will be reversed on the display for left and right directions.
Root causeNonconforming Material/Component
ActionKing Systems issued notification letter -Urgent Medical Device Recall to Domestic consignees notified via E-mail on 11/5/19 with attached letters and forms. Foreign consignees were notified according to local regulation. On January 24, 2020, King Systems issued a Press Release dated 1/20/2020. Customers are advised to take the following action: 1) Review your current inventory and place any affected products in quarantine immediately. 2) The affected products should NOT be used. 3) Please note any affected quantity of product(s) on the attached form and return it within 30 days from th
Quantity171 units
DistributionWorldwide distribution. US Nationwide, Australia, Belgium, Bolivia, Canada, Costa Rica, Germany, Hong Kong, India, Italy, Japan, Myanmar, Poland, Spain, and UK.