Recall Z-0867-2020 ยท LZP Aid, Surgical, Viscoelastic
Johnson & Johnson Surgical Vision Inc
Healon GV PRO 0.85 mL, Model (U.S.): 10-2400-14, Model (O.U.S.): 10-2705-14, 10-2715-14, 10-2725-14, and 10-2755-14
| Date | Jan 27, 2020 (initiated Dec 3, 2019) |
| Status | Terminated |
| Reason | Ophthalmic viscosurgical device may be difficult to remove from the eye, leading to increased postoperative intraocular pressure requiring additional intervention. Also, reports of clogging of phacoemulsion equipment tubing, which may lead to delay in the procedure or ocular injury. |
| Root cause | Labeling False and Misleading |
| Action | Customers were asked to immediately take the following actions: 1. Compare your inventory against the affected product list. 2. Discontinue using and remove from your inventory all affected lots. 3. Even if you have no inventory, please complete the Customer Reply Form. If you have inventory of any of the affected OVDs products, please complete the Customer Reply Form, noting the lot numbers of the OVDs, and use the shipping label provided to return the product. Returned product will NOT require refrigerated shipping. This notice should be shared with anyone who needs to be aware within your |
| Distribution | U.S.: CA, PA, NJ, GA, IL, MO, CO, TX, FL, AK, NY, HI, UT, LA, OH, IN, MA, NC, AZ, MI, NH, AL, TN, WA, MD, OR, VA, MS, MN, KY, AR, NV, WI, WV, SC, DE, ME, CT, ID, NM, VT, IA, OK, MT, NE, DC. OUS: DE, GB, DK, IT, NL, TR, AT, CH, FR, GR, ES, PT, FO, BE, GP, ZA, IE, HR, HU, IS, NO, SE, FI |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.