ClearanceBrief
Recall Z-0836-2020 ยท DZE Implant, Endosseous, Root-Form

Paltop Advanced Dental Solutions Ltd

Paltop Dynamic Dental Implant Dia 3.25mm L 13.0 mm, Catalog Number 21-70020

DateNov 18, 2019
StatusTerminated
ReasonSome of the implants were packaged with a Standard Platform Cover Screw instead of a Narrow Platform Cover Screw. The Standard Platform Cover Screw will not fit in the Narrow Platform implant.
Root causeMixed-up of materials/components
ActionUrgent: Medical Device Recall notification letters dated 11/18/19 were sent to customers. Please discontinue use. Attached is a Reply Communication form that should be used to let us know whether the product has been used or will be returned. Once we receive that Form, a Customer Service agent will reach out to you with return instructions and provide a label if necessary. Paltop Advanced Dental will replace the implants that are in stock.
Quantity67
DistributionThe products were distributed to the following US states: CA, DC, FL, GA, MA, MD, NJ, NY, SC, TX, and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130462PALTOP NARROW IMPLANT, PALTOP SHORT IMPLANT
Paltop Advanced Dental Solutions, Ltd., Aug 27, 2013