Recall Z-0693-2020 ยท DTM Filter, Blood, Cardiopulmonary Bypass, Arterial Line

Medtronic Perfusion Systems

Medtronic Affinity, 38um Arterial Filter, Catalog number REF 351

DateNov 14, 2019
StatusTerminated
ReasonMedtronic has identified an out of-specification condition exhibiting excess plastic (flash) in the arterial filter directly above the outlet port.
Root causeProcess control
ActionBeginning 14-Nov-2019, Medtronic initiated an immediate Urgent Medical Device Recall via verbal communication to one (1) customer that was shipped the affected product according to Medtronic records. The customer was asked to identify and quarantine any unused affected product. The immediate communication was followed by a written urgent medical device recall notification beginning on 26-Nov-2019 to communicate the issue to the one (1) customer that received affected product. The consignee was asked to confirm receipt of urgent medical device recall notification via a Customer Confirmation Ce
Quantity24 units
DistributionCA
FDA recordOpen on fda.gov