Recall Z-0715-2020 ยท DYG Catheter, Flow Directed

Arrow International Inc

Arrow Berman Angiographic Catheter 50cc Item Number: AI-07130

DateNov 15, 2019
StatusTerminated
ReasonPotential leak in the balloon extension line. If a leak was present and went undetected during pre-use testing as outlined in the IFU, this would cause the balloon to deflate slowly following placement. If the leak went undetected by attending clinicians during use, it could result in contrast injection into the vascular or cardiac tissue which could result in serious injury or death
Root causeProcess control
ActionArrow International issued letters on 11/15/19 stating reason for recall, health risk, and action to take: immediately discontinue use and quarantine any products with the product codes and lot numbers listed above. 2. To acknowledge receipt of this Urgent Medical Device Recall Notice or to return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Au
Quantity71 units
DistributionU.S. Nationwide, Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K892530CATHETERS FOR BALLOON OCCLUSION FEMORAL ANGIOGRAPH
Arrow Intl., Inc., Sep 26, 1989