Recall Z-0726-2020 ยท DSI Detector And Alarm, Arrhythmia

GE Healthcare, LLC

CIC Pro Clinical Information Center Central Station - Product Usage: The intended use is to provide clinicians with adult, pediatric and neonatal patient data in a centralized location within a hospital or clinical environment. CIC Pro Clinical Information Center central station is intended to collect information from a network and display this data. This data includes physiological, patient demog

DateNov 12, 2019
StatusTerminated
ReasonWhen connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) version 1 and Central Information Center (CIC) systems were identified to have vulnerabilities to a cyber-attack.
Root causeSoftware in the Use Environment
ActionThe firm disseminated the notices by mail on 11/12/2019. The letter identified the security vulnerability of certain GE Central Stations and ApexPro Telemetry servers and provided additional best practices, ensuring the following: 1. MC and IX Networks are isolated; 2. MC and IX Router/Firewalls block incoming traffic, as applicable; 3. Restricted physical access to Central Stations, Telemetry Servers, MC network and IX network; 4. Default passwords are changed as applicable; and 5. Password management best practices are adhered to Ensuring the networks are properly configured and isolated
Quantity41848 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032370CIC PRO CLINICAL INFORMATION CENTER CENTRAL STATION
Ge Medical Systems Information Technologies, Aug 13, 2003
K053356MODIFICATION TO: CIC PRO CLINICAL INFORMATION CENTER CENTRAL STATION
Ge Medical Systems Information Technologies, Apr 19, 2006
K062976MODIFICATION TO CIC PRO CLINICAL INFORMATION CENTER
Ge Medical Systems Information Technologies, Mar 16, 2007