ClearanceBrief
Recall Z-0802-2020 ยท CCW Laryngoscope, Rigid

Flexicare Medical Ltd.

BriteBlade Pro Single-se Fiber Optic Miller 2, CE 0120, Do not Re-use, Rx Only, REF, 040-722U, Sterile EO, Qty 10, (01) 15055788722783 The device is intended for direct vision laryngoscopy to obtain an illuminated view of the vocal cords and the glottis by insertion into the oral cavity to manipulate the tongue and/or the epiglottis.

DateJan 13, 2020 (initiated Nov 27, 2019)
StatusTerminated
ReasonThe firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.
Root causeNonconforming Material/Component
ActionOn 11/27/2019, the firm "Flexicare" sent a "Urgent: medical Device Recall" letter via USPS Priority mail. The letter informed consignees of the recall. In addition to informing consignees of the recall the letter asked consignees to take the following actions: Complete the acknowledgment and receipt form and return this form to Flexicare Inc. On receipt of the form you will be contacted by a representative of Flexicare Inc. to make all necessary further arrangements. The form asked the following of the consignee: -Acknowledge that you have read and understand the recall instructions provide
Quantity1,024,890 devices
DistributionUS: Nationwide OUS: Australia, Canada, China, Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.