Recall Z-0925-2020 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

da Vinci Xi Surgical System Surgical System, IS4000, 00886874114216 - Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.

DateJul 26, 2019
StatusTerminated
ReasonThe firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience either vision loss or non-recoverable errors 307 or 319, preventing further use of the system.
Root causeComponent design/selection
ActionOn July 26, 2019, Intuitive sent out a "Urgent Medical Device Correction" letter to affected consignees via FedEx, the Intuitive Customer Portal, and via Email. In addition, to informing their consignees about the recall, the letter also asked consignees to take the following actions: 1. If necessary, please follow the System Troubleshooting section in your specific system user manual for guidance on how to proceed in the scenario where vision loss or a non-recoverable error is experienced, including contacting Intuitive Technical Support. In addition: 2. Forward this letter to your Risk Ma
Quantity123 surgical systems
DistributionWorldwide distribution - US Nationwide distribution in the states of Alabama, Alaska, Arizona, California, Colorado, Connecticut, Florida, Hawaii, Idaho, Illinois, Iowa, Kansas, Kentucky, Louisiana, Massachusetts, Michigan, Minnesota, Missouri, Montana, New Jersey, New Mexico, New York, North Caroli
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131861DA VINCI SURGICAL SYSTEM, ENDOWRIST INSTRUMENTS AND ACCESSORIES
Intuitive Surgical, Inc., Mar 28, 2014