Recall Z-0670-2020 ยท MUJ System, Planning, Radiation Therapy Treatment

Elekta Inc

Monaco Radiation Treatment Planning System (RTP) System

DateNov 22, 2019
StatusOpen, Classified
ReasonMonaco is using the incorrect energy when optimizing and calculating dose.
Root causeSoftware Design Change
ActionAn Important Field Safety Notification letter is being emailed to affected customers beginning 11/22/2019. The letter identifies affected product, provides a description of the problem, and recommends user not use multiple energies within the same plan until a software fix has been applied to the device. A response form was asked to be returned. Affected customers are advised to contact local Elekta service teams for questions.
Quantity76 (7 US, 69 OUS)
DistributionIA, IN, NY, OH, TX, WI. Internationally distributed to: Australia, Canada, China, Denmark, Germany, Greece, Hong Kong, Italy, Netherlands, Sweden, Switzerland, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183037Monaco RTP System
Elekta, Inc., Dec 4, 2018
K190178Monaco RTP System
Elekta, Inc., Mar 29, 2019