Recall Z-0696-2020 ยท GFJ Mallet, Surgical, General & Plastic Surgery

Smith & Nephew, Inc.

smith&nephew PROFIX Mallet, 750 GRAMS, REF 71512452 - Product Usage: The PROFIX Mallet is a reusable surgical instrument used to tap instruments or implants during knee or hip surgery.

DateNov 11, 2019
StatusTerminated
ReasonThere is a a potential failure mode associated with the use of the mallet, wherein the poly tip of the mallet may become loose and/or come off during use. If the potential failure mode occurs, particles generated in the poly thread potentially could enter the surgical field or a delay in surgery could occur if the tip disengages during the use of the mallet.
Root causeDevice Design
ActionThe firm disseminated a medical device field correction notice on 11/11/2019 by letter. The letter explained the failure and requested the following actions be taken by hospital representatives: 1. Please check your inventory and complete the attached Inventory Correction Certification Form. 2. If you have the PROFIX 750 Gram Mallet, please maintain awareness of this notice. 3. When preparing for cleaning, disassemble the poly tip from the mallet and inspect disassembled components for damage or wear. 4. After cleaning, inspect disassembled components and ensure the components can be re-assem
Quantity1371 devices
DistributionWorldwide distribution - US Nationwide distribution and countries of Belgium, Canada, China, Columbia, Costa Rica, Finland, France, Germany, Great Britain, Hong Kong, Italy, Netherlands, Norway, Panama, Poland, Switzerland, United Arab Emirates, Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.