Recall Z-0730-2020 ยท NIP Stent, Superficial Femoral Artery

Cardinal Health Inc.

Cordis S.M.A.R.T. CONTROL /S.M.A.R.T. Vascular Stent System. The S.M.A.R.T. CONTROL and S.M.A.R.T. Vascular Stent Systems consist of a self-expanding stent made of Nitinol (nickel-titanium alloy) material that is pre-mounted on an over-the-wire delivery system. The stents contain tantalum markers at each end. The S.M.A.R.T. CONTROL and S.M.A.R.T. Vascular Stents are designed to open to a pr

DateNov 14, 2019
StatusOpen, Classified
ReasonCordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe".
Root causeUnder Investigation by firm
ActionCustomers were notified with a Medical Device Correction letter. The letter was sent via overnight delivery beginning 11/14/2019. The letter identified affected product, provided a description of the problem, and asked that a copy of the letter be retained where the product is stored. Customer were asked to completed and return an Acknowledgement Form. Questions can be directed to the Cordis Endovascular Medical Information Line 877-DEVICE5 (877-338-4235) Monday through Friday 9:00 am - 5:00 pm EST.
Quantity267,247 units total (262,641 OUS, 4,606 US)
DistributionWorldwide distribution - US Nationwide including Puerto Rico and countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaiian, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Repulic, Denmark, Dominican Republic, Ecua
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.