Recall Z-0729-2020 ยท NIM Stent, Carotid
Cardinal Health Inc.
Cordis Precise Nitinol Stent System. The Cordis PRECISE Nitinol Stent System devices are designed to deliver a flexible, self-expanding endoluminal stent to the carotid vasculature via over-the-wire (OTW) 5.5F or 6F sheathed delivery systems. The stent is cut from a solid nitinol tube into a fine mesh ( Z configuration) design. The delivery systems consist mainly of an inner shaft and an out
| Date | Nov 14, 2019 |
| Status | Open, Classified |
| Reason | Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe". |
| Root cause | Under Investigation by firm |
| Action | Customers were notified with a Medical Device Correction letter. The letter was sent via overnight delivery beginning 11/14/2019. The letter identified affected product, provided a description of the problem, and asked that a copy of the letter be retained where the product is stored. Customer were asked to completed and return an Acknowledgement Form. Questions can be directed to the Cordis Endovascular Medical Information Line 877-DEVICE5 (877-338-4235) Monday through Friday 9:00 am - 5:00 pm EST. |
| Quantity | 267,247 units total (262,641 OUS, 4,606 US) |
| Distribution | Worldwide distribution - US Nationwide including Puerto Rico and countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaiian, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Repulic, Denmark, Dominican Republic, Ecua |
| FDA record | Open on fda.gov |