Recall Z-0701-2020 ยท FTR Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled

Allergan PLC

Natrelle Inspira Sillicone-Filled Breast Implants for the following Styles and Sizes: 7607 SSM, 7608 SSF, 7609 SSX, 7513 SRL, 7514 SRLP, 7515 SRM, 7516 SRF, 7517 SRX, 7626 SCL, 7627 SCLP, 7628 SCM, 7629 SCF, 7630 SCX: 7607 SSM/UDI: Natrelle Inspira SSM 140cc BI (US) 10888628032835, Natrelle Inspira SSM 175cc BI (US) 10888628032842, Natrelle Inspira SSM 195cc BI (US) 10888628032859, Natrelle Inspi

DateNov 6, 2019
StatusTerminated
ReasonThe firm has became aware that 2D barcodes on breast implants and sizers may be unreadable by GS-1 configured scanners. These 2D barcodes contain a group separators between fixed-length element string in the UDI data chain of the 2D barcode, which is not formatted to the GS1 standard.
Root causeError in labeling
ActionAllergan sent out the field notification to consignees via email on 11/06/19. For consignees whom email was unavailable, a field notification letter was sent via overnight mail with tracking to all consignees.In addition to informing them about the correction, the notification requested consignees do the following: 1. You may continue to use these products. If your facility uses GS-1 configured scanners and the 2D barcode cannot be scanned, the UDI information can be entered manually by using the human readable information that appears next to the 2D barcode.: 1. To enter the UDI, For the
DistributionUS:AL, AR,AZ,CA,CO,DC,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,WA,WI,WY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.