Recall Z-0700-2020 ยท FTR Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled

Allergan PLC

Natrelle Silicone Filled Breast Implants for the following Styles and Sizes: Style 10, Style 15, Style 20, Style 40, Style 45 Style 10/UDI: NATRELLE Style 10-180 Gel Round Mod 10888628001206, NATRELLE Style 10-210 Gel Round Mod 10888628001213, NATRELLE Style 10-240 Gel Round Mod 10888628001220, NATRELLE Style 10-270 Gel Round Mod 10888628001237, NATRELLE Style 10-300 Gel Round Mod 10888628001244,

DateNov 6, 2019
StatusTerminated
ReasonThe firm has became aware that 2D barcodes on breast implants and sizers may be unreadable by GS-1 configured scanners. These 2D barcodes contain a group separators between fixed-length element string in the UDI data chain of the 2D barcode, which is not formatted to the GS1 standard.
Root causeError in labeling
ActionAllergan sent out the field notification to consignees via email on 11/06/19. For consignees whom email was unavailable, a field notification letter was sent via overnight mail with tracking to all consignees.In addition to informing them about the correction, the notification requested consignees do the following: 1. You may continue to use these products. If your facility uses GS-1 configured scanners and the 2D barcode cannot be scanned, the UDI information can be entered manually by using the human readable information that appears next to the 2D barcode.: 1. To enter the UDI, For the
DistributionUS:AL, AR,AZ,CA,CO,DC,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,WA,WI,WY