ClearanceBrief
Recall Z-0699-2020 ยท FWM Prosthesis, Breast, Inflatable, Internal, Saline

Allergan PLC

Natrelle Saline-Filled Breast Implant, for the following Styles and sizes, Style LP, Style 68 HP, Style 68 MP, Style LP/UDI: NATRELLE Style 68LP-125 Saline Round Low 10888628000513, NATRELLE Style 68LP-150 Saline Round Low 10888628000520, NATRELLE Style 68LP-175 Saline Round Low 10888628000537, NATRELLE Style 68LP-200 Saline Round Low 10888628000544, NATRELLE Style 68LP-225 Saline Round Low 1088

DateNov 6, 2019
StatusTerminated
ReasonThe firm has became aware that 2D barcodes on breast implants and sizers may be unreadable by GS-1 configured scanners. These 2D barcodes contain a group separators between fixed-length element string in the UDI data chain of the 2D barcode, which is not formatted to the GS1 standard.
Root causeError in labeling
ActionAllergan sent out the field notification to consignees via email on 11/06/19. For consignees whom email was unavailable, a field notification letter was sent via overnight mail with tracking to all consignees.In addition to informing them about the correction, the notification requested consignees do the following: 1. You may continue to use these products. If your facility uses GS-1 configured scanners and the 2D barcode cannot be scanned, the UDI information can be entered manually by using the human readable information that appears next to the 2D barcode.: 1. To enter the UDI, For the
DistributionUS:AL, AR,AZ,CA,CO,DC,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,WA,WI,WY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.