Recall Z-0720-2020 ยท GXI Probe, Radiofrequency Lesion

Abbott Medical

CLASSIC Radiofrequency Cannula - Curved, 10 cm, SMK Compatible Electrode, REF C-1005-S - Product Usage: Disposable Radiofrequency Cannulae are provided in either straight, curved, or curved blunt geometries with a stylet inserted in the cannula. Cannulae are insulated except for the active tip. The active tip determines the size of the radiofrequency lesion.

DateNov 11, 2019
StatusTerminated
ReasonA manufacturing error caused a single lot of 10mm active tip length Cannulas to be labeled as 5mm active tip length cannulas.
Root causeProcess control
ActionThe firm initiated the recall by hand delivered letter on 11/11/2019. The letter explained the issue and requested the return of the recalled product.
Quantity470 devices
DistributionWorldwide distribution - US Nationwide and country of Bermuda.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090955SMITH & NEPHEW RF CANNULATE
Smith & Nephew, Inc., Jul 16, 2009