Recall Z-0664-2020 ยท GXD Generator, Lesion, Radiofrequency

Abbott Medical

Radiofrequency Grounding Pad, REF RF-DGP-L

DateOct 14, 2019
StatusTerminated
ReasonAffected lots were manufactured with the protective release liner in an incorrect orientation. When attempting to remove this disposable release liner, a clear film may be left on the grounding pad which may interfere with patient skin contact and lead to uneven heating.
Root causeProcess control
ActionThe firm initiated the recall by letter hand delivered by an Abbott field representative. The letter requested that consignee cease use of the affected lot and return the product with the aid of a field representative.
Quantity8940 devices
Distributionnationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111576NT 2000 LESIONING GENERATOR
Neuro Therm, Inc., Sep 20, 2011