Recall Z-0723-2020 ยท GWQ Full-Montage Standard Electroencephalograph

LIFELINES NEURO COMPANY

ElectroTek Temporal Sensor Cable Gold - Long Length, Single Length - Product Usage: The electroencephalographic system ElectroTek is intended to acquire, display, and store the electrical activity of a patient s brain obtained by placing electrodes on the patient s scalp. The ElectroTek is intended to be used for such studies as electroencephalogram (EEG) and video EEG recording to aid in diagnosi

DateOct 11, 2019
StatusTerminated
ReasonPotential for one of the four temporal electrode cables used in with the ElectroTek EEG system having incorrect placement labels which could lead to incorrect placement of electrode for an EEG study.
Root causeComponent change control
ActionOn October 14, 2019 the firm distributed Urgent Medical Device Recall Communication letters by overnight mail to the consumer/user accounts with replacement cables stating: Please find enclosed six packages of orange Temporal EEG electrode cables for use with your Trackit M system to replace the incorrectly labelled orange Temporal EEG electrode cables. Please discontinue using the orange Temporal Electrode cables that were provided with your Trackit M kit at delivery. Please return all of the orange Temporal Electrode Cables originally provided with the kit to our attention in the postage-pa
Quantity12
DistributionUS Nationwide distribution in the states of TX, TN, NC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170441ElectroTek
Mobilemedtek, Jun 5, 2017