Recall Z-0843-2020 ยท KQO Automated Urinalysis System

Beckman Coulter Inc.

iChem VELOCITY Urine Chemistry Analyzer

DateNov 7, 2019
StatusTerminated
ReasonBeckman Coulter has become aware of an issue affecting Specific Gravity chemistry settings for the iChemVELOCITY Urine Chemistry System that prevents the trigger of a chemistry system flag S3 and may result in erroneous Specific Gravity (SG) results.
Root causeError in labeling
ActionThe firm, Beckman Coulter, sent an "URGENT MEDICAL DEVICE RECALL" letters on 11/07/2019. The letter explained the problem and how to recognize if the S3 flag has been disabled in the instrument. The letter further instructs the following: - Review results to determine if samples report an SG result of >1.060 or an SG result of S3 - If the SG result is >1.060 or S3, test samples for SG using an alternate method* determined by your laboratory - If persistent SG results are >1.060 or persistent S3 results are seen, contact your local Beckman Coulter Representative If you have any questions r
Quantity4314 units
DistributionWorldwide Distribution: US (nationwide) including Puerto Rico and countries of: Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Ecuador, France,